Smart Cleaner® is a unique tool for cleaning blockages in the turbine and quick coupling through the use of a helical pushing movement of an injector piston.

Smart Cleaner® serves two-purposes:

  1. Removal of blockages caused by micro particles carried in the water supply and inert biofilm residue left in the turbine.
  2. Prevention of initial residue/particle builds up by using the smart cleaner with hot water at two days interval.

 
(a) Fit the connector of the Smart Cleaner® to the turbine / quick coupling.
(b) Undo Smart Cleaner® threaded handle.
(c) Remove Smart Cleaner® piston.
(d) Half-fill Smart Cleaner® with hot water
(e) Return piston to Smart Cleaner®, push it inside with normal hand pressure.
(f) If not successful or resistance is too great for normal hand pressure, fit the threaded handle and screw it down to the stop.
 
 
“Smart Cupper”® is necessary when, in a multi hole turbine, only one hole is blocked.
1. Take the bore cup off the case and Connect it to the turbine.
2. Adjust the hole of the bore cup exactly above the blocked hole in the turbine.
3. Connect the turbine with the case clump and close the screw.
4. Connect the turbine to the “Smart Cleaner”® and run a routine cleaning.
 
“Smart connect”® is necessary for contra angle handpieces (E Type).
1. Connect the smart cleaner® to the smart connect®.
2. Connect contraengle handpiece to the smart connect.
3. Run a routine use – as shown in the “Smart Cleaner”® instructions.
 
“Smart tube”® is necessary when the turbine tube, in the dental unit is blocked.
1. Disconnect the tube from the dental unit.
2. Connect the pin side of the “smart tube”® to the dental unit tube.
3. Connect the other side of the smart tube to the smart cleaner ®.
4. Run a routine use - as shown in the smart cleaner instruction®.